跳到主要导航 跳到搜索 跳到主要内容

The study of PVP/Pd/IrO2 modified sensor for amperometric determination of sulfur dioxide

  • East China Normal University
  • Gifu University

科研成果: 期刊稿件文章同行评审

摘要

A novel electrochemical sensor for the detection of sulfur dioxide in both gas and solution is described. The chemically modified electrode was constructed by polymerizing (4-VP), palladium and iridium oxide (PVP/Pd/IrO2) onto a platinum microelectrode which exhibits excellent catalytic activity toward sulfite with an oxidation potential of + 0.50 V. The SO2 gas sensor is based on the PVP/Pd/IrO2 modified electrode as working electrode, Ag/AgCl electrode as reference electrode, Pt electrode as counter electrode and a porous film, which is in direct contact with the gas-containing atmosphere. The effect of different internal electrolyte solutions of hydrochloric acid, sulfuric acid, phosphates buffer solution, mixed solution of dimethyl sulfoxide and sulfuric acid to the determination of SO2 was also studied. The sensor was found to have a high current sensitivity, a short response time and a good reproducibility for the detection of SO2. It has good potential to be used in the field of environmental monitoring and controlling.

源语言英语
页(从-至)241-247
页数7
期刊Talanta
55
2
DOI
出版状态已出版 - 30 8月 2001

学术指纹

探究 'The study of PVP/Pd/IrO2 modified sensor for amperometric determination of sulfur dioxide' 的科研主题。它们共同构成独一无二的学术指纹。

引用此