跳到主要导航 跳到搜索 跳到主要内容

Study of an amperometric glucose sensor based on Pd-Ni/SiNW electrode

  • Shichao Hui
  • , Jian Zhang*
  • , Xuejiao Chen
  • , Huhua Xu
  • , Dianfei Ma
  • , Yanli Liu
  • , Bairui Tao
  • *此作品的通讯作者
  • East China Normal University

科研成果: 期刊稿件文章同行评审

摘要

An amperometric glucose sensor based on Pd-Ni/SiNW electrode has been investigated. The silicon nanowire (SiNW) electrodes were first fabricated by chemical etching, and then nickel and palladium particles were deposited onto the surfaces of SiNWs via electroless co-plating technique followed by annealing in nitrogen atmosphere at 350 °C for 300 s. The morphology of Pd-Ni/SiNW electrode was characterized by scanning electron microscope (SEM) and X-ray diffraction (XRD). The sensor performance was characterized by cyclic voltammetry (CV) and fixed potential amperometry techniques. In 0.1 M KOH alkaline medium with different glucose concentrations, the sensor shows an excellent sensitivity of 190.72 μA mM-1 cm-2 with the detection limit (S/N ratio = 3) of 2.88 μM. And it also exhibits superior anti-interference properties to the species including ascorbic acid (AA), uric acid (UA) and 4-acetamidophenol (AP). All results demonstrate that this Pd-Ni/SiNW electrode is a candidate with great potential for glucose detection.

源语言英语
页(从-至)592-597
页数6
期刊Sensors and Actuators B: Chemical
155
2
DOI
出版状态已出版 - 20 7月 2011

指纹

探究 'Study of an amperometric glucose sensor based on Pd-Ni/SiNW electrode' 的科研主题。它们共同构成独一无二的指纹。

引用此