Simultaneous monitoring of glucose, lactate, l-glutamate and hypoxanthine levels in rat striatum by a flow-injection enzyme electrode array system with in vivo microdialysis sampling

  • Fen Fen Zhang
  • , Qiao Wan
  • , Chen Xin Li
  • , Xiao Li Wang
  • , Zi Qiang Zhu
  • , Yue Zhong Xian
  • , Li Tong Jin*
  • , Katsunobu Yamamoto
  • *此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

A flow-injection enzyme electrode array system with in vivo microdialysis sampling is proposed for the simultaneous measurement of cerebral glucose, lactate, l-glutamate and hypoxanthine concentrations. The enzyme electrode array system was based on neutral red-doped silica (NRDS) nanoparticles as the electrocatalyst. These uniform NRDS nanoparticles (about 50 ± 3 nm) were prepared by a water-in-oil microemulsion method, and characterized by the transmission electron microscopy technique. The inside neutral red dopant maintained its high electron-activity, while the outside nano silica surface prevented neutral red from leaching out into the aqueous solutions and showed high biocompatibility. These nanoparticles were then mixed with the glucose oxidase, lactate oxidase, l-glutamate oxidase or xanthine oxidase, and immobilized on the four carbon electrode array, respectively. A thin Nafion ® film was coated on the enzyme layer to prevent interference such as from ascorbic acid and uric acid in the dialysate. The proposed flow-injection analysis with the NRDS-enzyme electrode array system enables simultaneous monitoring of trace levels of glucose, l-glutamate, lactate and hypoxanthine in rat striatum.

源语言英语
页(从-至)1-7
页数7
期刊Journal of Electroanalytical Chemistry
575
1
DOI
出版状态已出版 - 15 1月 2005

指纹

探究 'Simultaneous monitoring of glucose, lactate, l-glutamate and hypoxanthine levels in rat striatum by a flow-injection enzyme electrode array system with in vivo microdialysis sampling' 的科研主题。它们共同构成独一无二的指纹。

引用此