跳到主要导航 跳到搜索 跳到主要内容

Separation and determination of nitroaniline isomers by capillary zone electrophoresis with amperometric detection

  • Xifeng Guo
  • , Jin Lv
  • , Weidong Zhang
  • , Qingjiang Wang
  • , Pingang He*
  • , Yuzhi Fang
  • *此作品的通讯作者
  • East China Normal University

科研成果: 期刊稿件文章同行评审

摘要

In this paper, capillary zone electrophoresis with amperometric detection (CZE-AD) was firstly applied to the simultaneous separation and determination of nitroaniline positional isomers. The three analytes could be perfectly analyzed by using the buffer of extreme pH. The effects of several important factors were investigated to find optimum conditions. A carbon-disk electrode was used as working electrode. The optimal conditions were 40 mmol/L tartaric acid-sodium tartrate (pH 1.2) as running buffer, 17 kV as separation voltage and 1.10 V (versus saturated calomel reference electrode, SCE) as detection potential. Under the optimum conditions, o-, m- and p-nitroaniline were separated successfully and good linearity, reproducibility and recovery results were obtained. The detection limit for m-nitroaniline was as low as at 9.06 × 10-9 mol/L. This proposed method demonstrated long-term stability and reproducibility with relative standard deviations of less than 1.8% for migration time and 1.1% for peak areas. The utility of this method was demonstrated by monitoring dyestuff wastewater and the assay results were satisfactory.

源语言英语
页(从-至)121-125
页数5
期刊Talanta
69
1
DOI
出版状态已出版 - 15 3月 2006

指纹

探究 'Separation and determination of nitroaniline isomers by capillary zone electrophoresis with amperometric detection' 的科研主题。它们共同构成独一无二的指纹。