跳到主要导航 跳到搜索 跳到主要内容

Platinum particles-modified electrode for HPLC with pulsed amperometric detection of thiols in rat striatum

  • Xu Ni Cao
  • , Jin Hua Li
  • , Hai Hong Xu
  • , Li Lin
  • , Yue Zhong Xian
  • , Katsunobu Yamamoto
  • , Li Tong Jin*
  • *此作品的通讯作者
  • East China Normal University
  • BAS Co., Ltd.

科研成果: 期刊稿件文章同行评审

摘要

A new chemically modified electrode based on the immobilization of Pt particles is fabricated and exhibits electrocatalytic oxidation for L-cysteine (L-Cys), glutathione (GSH) and penicillamine (PEN) with relatively high sensitivity. It is also adaptable to HPLC for pulsed amperometric detection (PAD) of these thiols. PAD largely improves the detection sensitivity because the alternated polarizations can effectively clean and reactivate the electrode surface. It is shown that the peak currents of L-Cys, GSH and PEN are linear to their concentrations, with the calculated detection limit of 1.1 × 10-7, 1.8 × 10-7 and 3.8 × 10-7 mol L-1, respectively (S/N = 3). The method has been successfully applied to assess the contents of L-Cys and GSH in rat striatal microdialysates. The average contents of the two analytes in rat striatum are 2.6 × 10-6 and 2.8 × 10-6 mol L-1, respectively.

源语言英语
页(从-至)564-569
页数6
期刊Biomedical Chromatography
18
8
DOI
出版状态已出版 - 10月 2004