跳到主要导航 跳到搜索 跳到主要内容

Parallel interrogation of the chalcogenide-based micro-ring sensor array for photoacoustic tomography

  • Jingshun Pan
  • , Qiang Li
  • , Yaoming Feng
  • , Ruifeng Zhong
  • , Zhihao Fu
  • , Shuixian Yang
  • , Weiyuan Sun
  • , Bin Zhang
  • , Qi Sui
  • , Jun Chen
  • , Yuecheng Shen*
  • , Zhaohui Li*
  • *此作品的通讯作者
  • Sun Yat-Sen University
  • Southern Marine Science and Engineering Guangdong Laboratory - Guanzhou
  • South China Normal University

科研成果: 期刊稿件文章同行评审

摘要

Photoacoustic tomography (PAT), also known as optoacoustic tomography, is an attractive imaging modality that provides optical contrast with acoustic resolutions. Recent progress in the applications of PAT largely relies on the development and employment of ultrasound sensor arrays with many elements. Although on-chip optical ultrasound sensors have been demonstrated with high sensitivity, large bandwidth, and small size, PAT with on-chip optical ultrasound sensor arrays is rarely reported. In this work, we demonstrate PAT with a chalcogenide-based micro-ring sensor array containing 15 elements, while each element supports a bandwidth of 175 MHz (−6 dB) and a noise-equivalent pressure of 2.2 mPaHz−1/2. Moreover, by synthesizing a digital optical frequency comb (DOFC), we further develop an effective means of parallel interrogation to this sensor array. As a proof of concept, parallel interrogation with only one light source and one photoreceiver is demonstrated for PAT with this sensor array, providing images of fast-moving objects, leaf veins, and live zebrafish. The superior performance of the chalcogenide-based micro-ring sensor array and the effectiveness of the DOFC-enabled parallel interrogation offer great prospects for advancing applications in PAT.

源语言英语
文章编号3250
期刊Nature Communications
14
1
DOI
出版状态已出版 - 12月 2023
已对外发布

指纹

探究 'Parallel interrogation of the chalcogenide-based micro-ring sensor array for photoacoustic tomography' 的科研主题。它们共同构成独一无二的指纹。

引用此