跳到主要导航 跳到搜索 跳到主要内容

On-line sample preconcentration technique based on a dynamic pH junction in CE-amperometric detection for the analysis of biogenic amines in urine

  • Wanrong Tang
  • , Shuli Ge
  • , Fan Gao
  • , Guan Wang
  • , Qingjiang Wang*
  • , Pingang He
  • , Yuzhi Fang
  • *此作品的通讯作者
  • East China Normal University

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

It is difficult to detect biogenic amines (BAs) in biological fluids because of their very low concentrations. In this paper, we reported an on-line sample preconcentration method in CE-amperometric detection (CE-AD) based on a dynamic pH junction, and a concentration enhancement of approximately 100-fold was achieved compared with the classical CE-AD methods in the simultaneous analysis of six BAs in urine (dopamine, epinephrine, norepinephrine, tyramine, tryptamine, and serotonin). The technique is proposed based on the sharp pH change generated at the boundary between an acidic sample and the basic BGE zone. Under optimized conditions, all analytes were successfully focused and well separated within 20 min with high efficiency and sensitivity (LODs at S/N = 3 ranging from 5.34 to 68.3 nM). For the analysis of urine samples by this method, satisfactory recoveries were obtained without a complicated pretreatment step or derivatization process. Therefore, it is self-evident that this approach for the analysis of real biological samples has great potential in the future.

源语言英语
页(从-至)2041-2048
页数8
期刊Electrophoresis
卷34
期14
DOI
出版状态已出版 - 7月 2013

学术指纹

探究 'On-line sample preconcentration technique based on a dynamic pH junction in CE-amperometric detection for the analysis of biogenic amines in urine' 的科研主题。它们共同构成独一无二的学术指纹。

引用此