跳到主要导航 跳到搜索 跳到主要内容

Measurement of dissolved oxygen and its diffusivity in aerobic granules using a lithographically-fabricated microelectrode array

  • Shao Yang Liu
  • , You Peng Chen
  • , Fang Fang
  • , Juan Xu
  • , Guo Ping Sheng
  • , Han Qing Yu*
  • , Gang Liu
  • , Yang Chao Tian
  • *此作品的通讯作者
  • University of Science and Technology of China

科研成果: 期刊稿件文章同行评审

摘要

A novel gold microelectrode array (MEA) was manufactured with microfabrication techniques and applied on the measurement of dissolved oxygen profile in an aerobic granule. The MEA contained five gold microelectrodes, which had a good linear response to dissolved oxygen and typically had a lifetime of more than 10 days. Dissolved oxygen microprofiles near the surface of an aerobic granule were monitored with this MEA. Based on the measurements, an oxygen effective diffusivity in the upper 100 wm layer of the aerobic granule was estimated to be 1.19 x 10-9 m2/s. The experimental results demonstrate that the MEA was able to measure the DO levels in aerobic granules accurately and precisely and that the MEA could be used to determine constituents, profiles, and functions in situ in small spaces. Moreover, since the device shape and microelectrode arrangementwere all defined by photolithography, the proposed fabrication procedure was flexible and appropriate for fabrication of various types of MEAs.

源语言英语
页(从-至)1160-1165
页数6
期刊Environmental Science and Technology
43
4
DOI
出版状态已出版 - 15 2月 2009
已对外发布

指纹

探究 'Measurement of dissolved oxygen and its diffusivity in aerobic granules using a lithographically-fabricated microelectrode array' 的科研主题。它们共同构成独一无二的指纹。

引用此