Liquid chromatography - Electrochemical detector for the determination of glucose in rat brain combined with in vivo microdialysis

  • Guoyue Shi
  • , Jiaxing Lu
  • , Fang Xu
  • , Hongguo Zhou
  • , Litong Jin*
  • , Jiye Jin
  • *此作品的通讯作者
  • East China Normal University
  • Gifu University

科研成果: 期刊稿件文章同行评审

摘要

In this paper, microdialysis sampling was combined with liquid chromatography with electrochemical detection (LC-ECD) and glucose in rat brain was determined. A cobalt tetraaminophthalocyanine (CoTAPC)/glucose oxidase (GOD)/Nafion film-coated glassy carbon electrode was used to selectively detect the hydrogen peroxide generated enzymatically in the enzyme reactors, without any interference from electroactive species such as L-ascorbate, urate, proteins. The signal current was linearly related to the glucose concentration in the range of 1.0x10-6-5.0x10-3mol/l at +0.5V (versus Ag/AgCl) with a detection limit of 5.0x10-7mol/l. To measure glucose in rat brain, microdialysis sampling was employed. The dialysate from the microdialysis system was injected into system. The result was 0.74±0.10mmol/l (n=4). Copyright (C) 2000 Elsevier Science B.V.

源语言英语
页(从-至)131-136
页数6
期刊Analytica Chimica Acta
413
1-2
DOI
出版状态已出版 - 18 5月 2000

指纹