In Vivo Measurement of Taurine With Optimized Double-Quantum Filtering Magnetic Resonance Spectroscopy

  • Yufan Zhou
  • , Chenting Ye
  • , Jiaqiang Zhou
  • , Yang Cao
  • , Dingyi Lin
  • , Yefeng Yao
  • , Hong Zhang
  • , Min Wang*
  • *此作品的通讯作者
  • Key Laboratory for Biomedical Engineering of Ministry of Education
  • Department of Endocrinology
  • Zhejiang University
  • Department of Nuclear Medicine and PET Center
  • The Second Affiliated Hospital of Zhejiang University School of Medicine

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

Purpose: To enhance the accuracy and reproducibility for in vivo taurine quantification using the double-quantum filtering (DQF) technique. Methods: Simulations and phantom experiments were conducted at both 7 T and 3 T to determine the optimal parameter τ by calculating the taurine purity index among other neighboring metabolites. The frequency-selective read-pulse bandwidth and the dephasing/rephasing gradient moments were further adjusted to reduce residual water side band interference. Rat models of C6 gliomas were compared to normal brain for taurine level at 7 T. Metabolomics was conducted to validate the accuracy of taurine quantification. Human gray and white matter taurine differences were revealed at 3 T. Test–retest reliability was evaluated on humans. Results: Simulations and phantom experiments identified optimal τ for maximizing taurine signal purity as 102 ms at 7 T and 92 ms at 3 T. The best combinations of parameters for minimal water interference were determined at 3 T. Optimized DQF sequences yielded lower CRLBs. In rat brains, taurine concentration measured with optimized DQF sequences exhibited strong correlation with metabolomics, outperforming both conventional MRS and nonoptimized DQF sequences. Only the optimized method revealed significant taurine concentration differences between gray and white matter. Test–retest demonstrated good reproducibility in gray matter (ICC = 0.82) for the optimized method. Conclusion: The optimized DQF method enhances taurine detection accuracy and robustness in both animal and human studies and was for the first time applied in human brain taurine test at 3 T.

源语言英语
页(从-至)2058-2071
页数14
期刊Magnetic Resonance in Medicine
卷96
期5
DOI
出版状态已接受/待刊 - 2026

学术指纹

探究 'In Vivo Measurement of Taurine With Optimized Double-Quantum Filtering Magnetic Resonance Spectroscopy' 的科研主题。它们共同构成独一无二的学术指纹。

引用此