Improved sensing membrane immobilization for enhanced long-term stability of iodide ion-selective microelectrode

  • Yun Chen
  • , Jian Zhang
  • , Yi Ting Wang*
  • , Jing Wang
  • , Yan Jie Bao
  • , Wei Jie Du
  • , Jun Jun Li
  • , Hui Peng
  • , Jian Zhong Zhu
  • *此作品的通讯作者
  • East China Normal University
  • Nantong University

科研成果: 期刊稿件文章同行评审

摘要

A porous silicon (PS)-based iodide ion-selective microelectrode (I-ISME) was fabricated by using a microelectronic planar process and an electrochemical anodization etching technique. The apparent sensing area of the I-ISME is 100×100 μm2. The response time t95% is 8 and 12 s when the concentration change is from low to high and vice versa, respectively. The potentials are constant at pH 1.0-2.0. The calibration curve for the I-ISME is linear within a wide range of 5.0×10-7-1.0×10 -1 M. Its average slope during 12 months is 58.0 mV per decade, which is close to the Nernst response. The detection limit was found to be on the order of 1.0×10-7 M. The sequence of Hofmeister selectivity coefficient (Kpoti,j) of the I-ISME were I-Br- SCN- Cl- SO2-4 NO-3 . Good performances of the I-ISME are attributed to large specific surface area of PS and excellent adhesion between sensing carrier and the surface of PS.