摘要
A porous silicon (PS)-based iodide ion-selective microelectrode (I-ISME) was fabricated by using a microelectronic planar process and an electrochemical anodization etching technique. The apparent sensing area of the I-ISME is 100×100 μm2. The response time t95% is 8 and 12 s when the concentration change is from low to high and vice versa, respectively. The potentials are constant at pH 1.0-2.0. The calibration curve for the I-ISME is linear within a wide range of 5.0×10-7-1.0×10 -1 M. Its average slope during 12 months is 58.0 mV per decade, which is close to the Nernst response. The detection limit was found to be on the order of 1.0×10-7 M. The sequence of Hofmeister selectivity coefficient (Kpoti,j) of the I-ISME were I-Br- SCN- Cl- SO2-4 NO-3 . Good performances of the I-ISME are attributed to large specific surface area of PS and excellent adhesion between sensing carrier and the surface of PS.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 699-703 |
| 页数 | 5 |
| 期刊 | Nanoscience and Nanotechnology Letters |
| 卷 | 5 |
| 期 | 6 |
| DOI | |
| 出版状态 | 已出版 - 6月 2013 |
学术指纹