摘要
An amperometric biosensor for hydrogen peroxide (H2O2) has been constructed by immobilizing cytochrome c on an indium/tin oxide (ITO) electrode modified with a macroporous material. Cyclic voltammetry showed that the direct and quasi-reversible electron transfer of cytochrome c proceeds without the need for an electron mediator. A surface-controlled electron transfer process can be observed with an apparent heterogeneous electron-transfer rate constant (ks) of 29.2 s-1. The biosensor displays excellent electrocatalytic responses to the reduction of H2O2 to give amperometric responses that increase steadily with the concentration of H2O2 in the range from 5 μM to 2 mM. The detection limit is 0.61 μM at pH 7.4. The apparent Michaelis-Menten constant (Km) of the biosensor is 1.06 mM. This investigation not only provided a method for the direct electron transfer of cytochrome c on macroporous materials, but also established a feasible approach for durable and reliable detection of H2O2.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 87-95 |
| 页数 | 9 |
| 期刊 | Microchimica Acta |
| 卷 | 175 |
| 期 | 1-2 |
| DOI | |
| 出版状态 | 已出版 - 10月 2011 |
| 已对外发布 | 是 |
学术指纹
探究 'Electrochemistry and biosensing activity of cytochrome c immobilized in macroporous materials' 的科研主题。它们共同构成独一无二的学术指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver