跳到主要导航 跳到搜索 跳到主要内容

Determination of thiol compounds in rat striatum microdialysate by HPLC with a nanosized CoHCF-modified electrode

  • Meichuan Liu
  • , Ping Li
  • , Yuxiao Cheng
  • , Yuezhong Xian
  • , Chenglin Zhang
  • , Litong Jin*
  • *此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

A cobalt hexacyanoferrate (CoHCF) nanoparticle (size ca. 60 nm) chemically modified electrode (CME) was fabricated and the electrochemical behavior of thiols at this nanosized CoHCF CME was studied. In comparison with a bare glassy carbon (GC) electrode and with a general CoHCF CME which was electrode-posited in the traditional manner, the present nanosized CoHCF CME efficiently performed electrocatalytic oxidation for glutathione (GSH) and L-Cysteine (L-Cys) with relatively high sensitivity, outstanding stability, and long-life. Combined with high-performance liquid chromatography (HPLC), the nanosized CoHCF CME was used for electrochemical determination (ECD) of GSH and L-Cys. The peak currents were a linear function of concentrations in the range 2.0×10-7 to 2.0×10-4 mol L-1 for both GSH and L-Cys, with detection limits of 1.2×10-7 and 1.0×10-7 mol L-1, respectively. Coupled with microdialysis sampling, the HPLC ECD system has been successfully used to assess the GSH and L-Cys content of rat striatum.

源语言英语
页(从-至)742-750
页数9
期刊Analytical and Bioanalytical Chemistry
380
5-6
DOI
出版状态已出版 - 11月 2004

指纹

探究 'Determination of thiol compounds in rat striatum microdialysate by HPLC with a nanosized CoHCF-modified electrode' 的科研主题。它们共同构成独一无二的指纹。