Automated Ultrasound Measurement of the Inferior Vena Cava: An Animal Study

  • Jiangang Chen
  • , Jiawei Li
  • , Xin Ding
  • , Gaofeng Wei
  • , Xiaoting Wang
  • , Qingli Li*
  • *此作品的通讯作者

科研成果: 期刊稿件文章同行评审

摘要

Because of continuous movement and variation in diameter of the inferior vena cava (IVC) with respiration, the measurements on IVC are labor-intensive and with considerable inter-operator variations. Some computer-assisted techniques have been developed to track the movement of the IVC semi-automatically. However, existing methods predominantly rely on reference marker selection and require many manual inputs. In this study, we developed a cross-correlation (CC)-based method for automated IVC movement tracking and measurement, which requires minimal manual input and avoids manual selection of reference markers. Based on the CC method, two approaches, named direct and relative approaches, were used to calculate the maximum, minimum, and variation of the IVC diameter, and compared with the manual measurement. Fifty-four ultrasound cine-loops collected from nine pigs were tested. The results reveal that both the proposed approaches were well agreed with the manual measurement. The errors of the direct approach were less than 9%, while those of relative approach were as high as 26.7%. It is concluded that the proposed direct approach is superior for IVC diameter measurements, which can be comparable with manual counterpart, serving as an alternative to traditional IVC measurement.

源语言英语
页(从-至)148-158
页数11
期刊Ultrasonic Imaging
42
3
DOI
出版状态已出版 - 1 5月 2020