跳到主要导航 跳到搜索 跳到主要内容

Analysis of the blackbody-radiation shift in an ytterbium optical lattice clock

  • Yi Lin Xu
  • , Xin Ye Xu*
  • *此作品的通讯作者
  • East China Normal University

科研成果: 期刊稿件文章同行评审

摘要

We accurately evaluate the blackbody-radiation shift in a 171Yb optical lattice clock by utilizing temperature measurement and numerical simulation. In this work. three main radiation sources are considered for the blackbody-radiation shift, including the heated atomic oven, the warm vacuum chamber, and the room-temperature vacuum windows. The temperatures on the outer surface of the vacuum chamber are measured during the clock operation period by utilizing seven calibrated temperature sensors. Then we infer the temperature distribution inside the vacuum chamber by numerical simulation according to the measured temperatures. Furthermore, we simulate the temperature variation around the cold atoms while the environmental temperature is fluctuating. Finally, we obtain that the total blackbody-radiation shift is -1.289(7) Hz with an uncertainty of 1.25 × 10-17 for our 171Yb optical lattice clock. The presented method is quite suitable for accurately evaluating the blackbody-radiation shift of the optical lattice clock in the case of lacking the sensors inside the vacuum chamber.

源语言英语
文章编号103202
期刊Chinese Physics B
25
10
DOI
出版状态已出版 - 10月 2016
已对外发布

学术指纹

探究 'Analysis of the blackbody-radiation shift in an ytterbium optical lattice clock' 的科研主题。它们共同构成独一无二的学术指纹。

引用此