摘要
Herein, we present an electrochemophysiological microarray (ECPM) for real-time mapping and simultaneous quantification of chemical signals for multiple ions in the deep brain of a freely moving rat, in which microelectrode arrays were developed for direct determination of multiple ions using open-circuit potentiometry. Specific recognition ionophores were synthesized and optimized for determination of K+, Ca2+, Na+ and pH. A reference electrode was also developed to avoid interferences in the brain. The microarrays were successfully applied in real-time monitoring and quantification of ions in a live brain. The extra current-free potentiometry allowed mapping and biosensing of chemical signals, together with recording of electrical signals in the whole brain without cross-talk, for the first time. Furthermore, the ECPM provided a platform for real-time monitoring of the dynamic changes of multiple ions in the deep brain of freely moving rat during a seizure.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 10426-10430 |
| 页数 | 5 |
| 期刊 | Angewandte Chemie - International Edition |
| 卷 | 59 |
| 期 | 26 |
| DOI | |
| 出版状态 | 已出版 - 22 6月 2020 |
学术指纹
探究 'An Electrochemophysiological Microarray for Real-Time Monitoring and Quantification of Multiple Ions in the Brain of a Freely Moving Rat' 的科研主题。它们共同构成独一无二的学术指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver