跳到主要导航 跳到搜索 跳到主要内容

Amperometric determination of chemical oxygen demand with flow injection analysis using F-PbO2 modified electrode

  • Jiaqing Li
  • , Luoping Li
  • , Lei Zheng
  • , Yuezhong Xian
  • , Shiyun Ai
  • , Litong Jin*
  • *此作品的通讯作者
  • East China Normal University
  • Shandong Agricultural University

科研成果: 期刊稿件文章同行评审

摘要

An electrochemical detection system with an F-PbO2 modified electrode for flow injection analysis to determine chemical oxygen demand (COD) was proposed and experimentally validated. The measuring principle was based on the current response on the modified electrode which was proportional to the COD value. Under the optimized experiment condition, the linear range was 100-1200 mg l-1, and the detection limit was 15 mg l-1. This method was successfully applied to determine the COD in samples without pretreatment. It was characterized by short analysis time, simplicity, low environmental impact, a limited reagent consumption and easy automation. Additionally, the COD values obtained from the proposed and conventional methods agreed well as demonstrated by the high significant correlation between the two sets of COD values (R = 0.9985, n = 25).

源语言英语
页(从-至)199-204
页数6
期刊Analytica Chimica Acta
548
1-2
DOI
出版状态已出版 - 29 8月 2005

指纹

探究 'Amperometric determination of chemical oxygen demand with flow injection analysis using F-PbO2 modified electrode' 的科研主题。它们共同构成独一无二的指纹。