跳到主要导航 跳到搜索 跳到主要内容

All-solid-state K+ sensing array based on Au@polystyrene nanocomposites

  • Fan Wang*
  • , Yalan Wang
  • , Linbo Li
  • , Chenyang Yuan
  • , Fan Zhang
  • , Wanqing Zhang
  • , Tianyou Yang*
  • *此作品的通讯作者
  • Henan Institute of Science and Technology

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

All-solid-state ion selective electrodes (ASS-ISEs) are easy to miniaturize and array, meeting the needs of home sensing devices. However, ASS-ISEs still faces challenges in accuracy and stability due to basic potential changes caused by non-specific adsorption of charged background compositions and the complex electrode preparation steps. To this end, our group successfully subtracted the background signal by integrating a self-calibrating channel in the sensing array and simplified the electrode preparation steps by preparing multi-functional PS-Au nanocomposites. However, the uniformity and gold content of PS-Au nanocomposites are difficult to control, so Au@PS nanocomposites are prepared as sensor materials in this paper to further reduce the differences between batches of electrodes. K+ Au@PS sensing array can be obtained by directly dropping Au@PS nanocomposites on the screen-printed carbon electrodes (SPCEs), which shows a near Nernstian behavior in the range 1.0 × 10−3 M to 0.3 M and good reproducibility in real sample testing. The detection results by K+ Au@PS sensing array for K+ in human morning urine agreed well with that tested by ICP-AES, which make the K+-ASS-ISE suitable for home health monitoring. Graphical Abstract: (Figure presented.)

源语言英语
期刊论文编号624
期刊Microchimica Acta
卷191
期10
DOI
出版状态已出版 - 10月 2024

联合国可持续发展目标

此成果有助于实现下列可持续发展目标:

  1. 可持续发展目标 3 - 良好健康与福祉
    可持续发展目标 3 良好健康与福祉

学术指纹

探究 'All-solid-state K+ sensing array based on Au@polystyrene nanocomposites' 的科研主题。它们共同构成独一无二的学术指纹。

引用此