A carbon fiber microelectrode-based third-generation biosensor for superoxide anion

  • Yang Tian
  • , Lanqun Mao
  • , Takeyoshi Okajima
  • , Takeo Ohsaka*
  • *此作品的通讯作者
  • Institute of Science Tokyo
  • CAS - Institute of Chemistry

科研成果: 期刊稿件文章同行评审

摘要

Implantable and miniature carbon fiber microelectrode (CFME)-based third-generation biosensor for superoxide anion (O2.-) was fabricated for the first time. The CFME-based biosensor was constructed by electro-deposition of Au nanoparticles on the CFMEs and then modification of the Au nanoparticles by cysteine followed by immobilization of superoxide dismutase (SOD) on the electrodes. The direct electrochemistry of the SOD immobilized on the CFME-based electrodes was efficiently realized by electron transfer promoter - cysteine molecules confined on the Au nanoparticles deposited on the CFMEs. The CFME-based biosensors were demonstrated to possess striking analytical properties for O2.- determination, such as optional operation potentials, high selectivity and sensitivity as well as good stability. Along with the implantable capacity inherent in the CFMEs, these striking analytical properties of the CFME-based biosensors substantially make them potential for in vivo determination of O2.-.

源语言英语
页(从-至)557-564
页数8
期刊Biosensors and Bioelectronics
21
4
DOI
出版状态已出版 - 15 10月 2005