跳到主要导航 跳到搜索 跳到主要内容

采用自动纯音测听的骨导听阈评估

  • Jiehua Chen
  • , Jinqiu Sang
  • , Jie Wang*
  • , Xiaoya Wang
  • *此作品的通讯作者
  • Guangzhou University
  • Guangzhou Women and Children’s Medical Center

科研成果: 期刊稿件文章同行评审

摘要

A bone conduction automatic audiometry scheme with adaptive masking for the non-test ear was implemented, and a complete automated pure tone audiometry program based on personal terminals was developed. Automated pure tone audiometry was conducted on 20 normal hearing subjects, and the results were compared with those obtained using the AC40 clinical audiometer. The majority of estimated air conduction thresholds (96%) and bone conduction thresholds (90%) differed from the clinical audiometer results by less than 10 dB. Furthermore, in the Bland-Altman plots of air and bone conduction thresholds and the air-bone gap, the data largely fell within the 95% confidence interval. The average difference and mean absolute difference in bone conduction thresholds between the automated system and the clinical audiometer were −0.17 dB and 4.52 dB, respectively, both generally smaller than the differences reported in existing automated audiometry systems.

投稿的翻译标题Assessment of bone conduction thresholds using automated pure tone audiometry
源语言繁体中文
页(从-至)1232-1241
页数10
期刊Shengxue Xuebao/Acta Acustica
49
6
DOI
出版状态已出版 - 11月 2024