The study of PVP/Pd/IrO2 modified sensor for amperometric determination of sulfur dioxide

Research output: Contribution to journalArticlepeer-review

42 Scopus citations

Abstract

A novel electrochemical sensor for the detection of sulfur dioxide in both gas and solution is described. The chemically modified electrode was constructed by polymerizing (4-VP), palladium and iridium oxide (PVP/Pd/IrO2) onto a platinum microelectrode which exhibits excellent catalytic activity toward sulfite with an oxidation potential of + 0.50 V. The SO2 gas sensor is based on the PVP/Pd/IrO2 modified electrode as working electrode, Ag/AgCl electrode as reference electrode, Pt electrode as counter electrode and a porous film, which is in direct contact with the gas-containing atmosphere. The effect of different internal electrolyte solutions of hydrochloric acid, sulfuric acid, phosphates buffer solution, mixed solution of dimethyl sulfoxide and sulfuric acid to the determination of SO2 was also studied. The sensor was found to have a high current sensitivity, a short response time and a good reproducibility for the detection of SO2. It has good potential to be used in the field of environmental monitoring and controlling.

Original languageEnglish
Pages (from-to)241-247
Number of pages7
JournalTalanta
Volume55
Issue number2
DOIs
StatePublished - 30 Aug 2001

Keywords

  • Determination of SO
  • PVP/Pd/IrO modified electrode
  • SO gas sensor

Fingerprint