A NTC thermistor-based wearable body temperature sensor was designed. This paper described the design principles and realization method of the NTC-based body temperature sensor. In this paper the temperature measurement error sources of the body temperature sensor were analyzed in detail. The automatic measurement and calibration method of ADC error was given. The results showed that the measurement accuracy of calibrated body temperature sensor is better than ± 0.04 degrees C. The temperature sensor has high accuracy, small size and low power consumption advantages.

Original languageEnglish
Pages (from-to)395-399
Number of pages5
JournalZhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Volume39
Issue number6
StatePublished - 1 Nov 2015