Skip to main navigation Skip to search Skip to main content

An energy-dependent electro-thermal response model of CUORE cryogenic calorimeter

  • D. Adams
  • , C. Alduino
  • , K. Alfonso
  • , F. Avignone
  • , O. Azzolini
  • , G. Bari
  • , F. Bellini
  • , G. Benato
  • , M. Beretta
  • , M. Biassoni
  • , A. Branca
  • , C. Brofferio
  • , C. Bucci
  • , J. Camilleri
  • , A. Caminata
  • , A. Campani
  • , L. Canonica
  • , X. Cao
  • , S. Capelli
  • , C. Capelli
  • L. Cappelli, L. Cardani, P. Carniti, N. Casali, E. Celi, D. Chiesa, M. Clemenza, S. Copello, O. Cremonesi, R. Creswick, A. D'Addabbo, I. Dafinei, F. Del Corso, S. Dell'Oro, S. Di Domizio, S. Di Lorenzo, V. Dompè, D. Fang, G. Fantini, M. Faverzani, E. Ferri, F. Ferroni, E. Fiorini, M. Franceschi, S. Freedman, S. Fu, B. Fujikawa, S. Ghislandi, A. Giachero, A. Gianvecchio, L. Gironi, A. Giuliani, P. Gorla, C. Gotti, T. Gutierrez, K. Han, E. Hansen, K. Heeger, R. Huang, H. Huang, J. Johnston, G. Keppel, Yu G. Kolomensky, R. Kowalski, M. Li, R. Liu, L. Ma, Y. Ma, L. Marini, R. Maruyama, D. Mayer, Y. Mei, S. Morganti, T. Napolitano, M. Nastasi, J. Nikkel, C. Nones, E. Norman, A. Nucciotti, I. Nutini, T. O'Donnell, M. Olmi, J. Ouellet, S. Pagan, C. Pagliarone, L. Pagnanini, M. Pallavicini, L. Pattavina, M. Pavan, G. Pessina, V. Pettinacci, C. Pira, S. Pirro, S. Pozzi, E. Previtali, A. Puiu, S. Quitadamo, A. Ressa, C. Rosenfeld, S. Sangiorgio, B. Schmidt, N. Scielzo, V. Sharma, V. Singh, M. Sisti, D. Speller, P. Surukuchi, L. Taffarello, F. Terranova, C. Tomei, K. Vetter, M. Vignati, S. Wagaarachchi, B. Wang, B. Welliver, J. Wilson, K. Wilson, L. Winslow, S. Zimmermann, S. Zucchelli
  • University of South Carolina
  • University of California at Los Angeles
  • Virginia Polytechnic Institute and State University
  • National Institute for Nuclear Physics
  • University of Rome La Sapienza
  • University of California at Berkeley
  • University of Milan - Bicocca
  • University of Genoa
  • Massachusetts Institute of Technology
  • Fudan University
  • Lawrence Berkeley National Laboratory
  • Gran Sasso Science Institute
  • University of Bologna
  • Université Paris-Saclay
  • California Polytechnic State University, San Luis Obispo
  • Shanghai Jiao Tong University
  • Yale University
  • Johns Hopkins University
  • P.O. Box 808 L-453

Research output: Contribution to journal › Article › peer-review

Abstract

The Cryogenic Underground Observatory for Rare Events (CUORE) is the most sensitive experiment searching for neutrinoless double-beta decay (0νββ) in 130Te. CUORE uses a cryogenic array of 988 TeO2 calorimeters operated at ∼10 mK with a total mass of 741 kg. To further increase the sensitivity, the detector response must be well understood. Here, we present a non-linear thermal model for the CUORE experiment on a detector-by-detector basis. We have examined both equilibrium and dynamic electro-thermal models of detectors by numerically fitting non-linear differential equations to the detector data of a subset of CUORE channels which are well characterized and representative of all channels. We demonstrate that the hot-electron effect and electric-field dependence of resistance in NTD-Ge thermistors alone are inadequate to describe our detectors' energy-dependent pulse shapes. We introduce an empirical second-order correction factor in the exponential temperature dependence of the thermistor, which produces excellent agreement with energy-dependent pulse shape data up to 6 MeV. We also present a noise analysis using the fitted thermal parameters and show that the intrinsic thermal noise is negligible compared to the observed noise for our detectors.

Original languageEnglish
Article numberP11023
JournalJournal of Instrumentation
Volume17
Issue number11
DOIs
StatePublished - 1 Nov 2022
Externally publishedYes

Keywords

  • Cryogenic detectors
  • Detector modelling and simulations I (interaction of radiation with matter, interaction of photons with matter, interaction of hadrons with matter, etc)
  • Double-beta decay detectors

Fingerprint

Dive into the research topics of 'An energy-dependent electro-thermal response model of CUORE cryogenic calorimeter'. Together they form a unique fingerprint.

Cite this